MLS-12

CAD data download
(DXF)
*Under construction
CATALOGUE
data download
(PDF)
MLS-12


Outside dimensions
Specifications

Type name

Item

MLS-12--250

Measuring range 250mm
Supply voltage DC5V to 12V±5%
Current consumption 40mA or less (under no load)
Detection system Incremental
Stroke speed mm/sec 250
Wire tensile force 0.29N~0.59N(30~60gf)
0.39N to 0.78N (40 to 80gf)
Output pulse number
[Minimum resolution]
60 300 600 1500
[1.0mm] [0.2mm] [0.1mm] 0.04mm]
Output phase A,B phase
Output form Square wave
Output capacity Sink current: 20mA
Residual voltage 0.5V
or less (at 10mA)
Maximum response frequency
(response pulse number)
50kHz
Working ambient
temperature/humidity
0 to 50
RH95% no dewing
Storing ambient temperature -20 to 80
Vibration resistance Durability 55Hz, double amplitude 1.5mm
2 hours each in X, Y, and Z directions
Impact resistance Durability 500m/s2 (about 50G)
3 times each in X, Y, and Z directions
Cable Outside diameter 3 4-core vinyl wire Insulated shield cable
Mass 60g


Specifications/micro linear scale
Type name

MLS-12-01-250

MLS-12-001-
250

Measuring range mm 250
Output pulse/1mm 10 100
Counter accuracy pulse ±1 ±1
Minimum reading 0.1 0.01
Power AC100V 50/60Hz 180mA max
Counter mode 10 Decimal
Count indication 0 to ±999999.9 0 to ±99999.99
Reset Reset switch provided *External signal resetting possible
External output *BCD parallel output
Mass 700g
Power cord 2m with plug
Note: For counter specifications, see DC Series. *indicates option. 

Output circuit diagram

Output waveform

BACK